Home > Services and Facilities > Reliability Engineering

Climatic Test Chamber

Manufacturer Feutron GmbH 3533/15 with simulation of natural Sunlight, IR, UV
Model 3533/15 with simulation of natural Sunlight, IR, UV
Application Simulation of natural Sunlight, IR, UV, Reliability testing
Standard Compliance IEC 68-2-14 Test Nb, IEC 68-2-30 Test Db, DIN 53384 DIN 50011, DIN IEC 60068-1, DIN EN 60068-2-1, DIN EN 60068-2-2, DIN 60068-2-3, DIN IEC 68 Teil 2-5, DIN IEC 68 Teil 2-9, DIN IEC 68-3-1
MIL-STD 810 C Meth. 502.1, MIL-StD 810 D Meth 501.2, MIL- STD 331A Teil 112.1, MIL-STD 202 F Meth 106 E, MIL-STD 810 D Tab 501.2-II, MIL-STD 810D Meth 507.2

Main Specification

Temperature inside Useful Test-room with Sunlight

Temperature Range -75 ... +180°C; -20...100°C
Temperature Range for Climatic Conditions +10 ... +95°C; 20...60°C
Temperature Deviation over Time ≤±0,2 ... ±0,5°C; ≤±0,5...1,0
Speed of Temperature Changes during Cooling Process +20°C to -70°C in <= 60min up to 3°C /min in the average
Speed of Temperature Changes during Heating Process -70°C to +180°C in <= 60min up to 4.5°C /min in the average
 

Humidity inside the Test-room with Sunlight

Total Range of Dew-point Temperature +5 ... +80°C
Range of Relative Humidity 5 ~ 98 %R.H: 10 ~ 80%R.H (with sunlight).
Deviation of Relative Humidity over Time ≤±3 %; ≤±4 %
 

Test-room

Test-room dimensions (W x H x D) 720 x 860 x 620 (mm)
Useful Test-room 400 liter
 

Wavelengths Covered

With Filter D 65 Terrestrial sunlight spectral region 340...400nm
With Filter AMO Extraterrestril sunlight
 

Transmitted Energy

At a Distance of 50 cm 650W / m2 max.
At a Distance of 100cm 220W/ m2 max.