Home > Services and Facilities > Reliability Engineering

Thermal Shock Test Chamber

Manufacturer Feutron GmbH 3416/15
Model 3416/15
Application Reliability testing, Accelerated life testing, Environmental Stress screening
Standard Compliance DIN 40046. Bl. 14, IEC 68 Teil 2-14 Na/Nb, IPC-TM-650 MIL STD 883 D Methode 1010.7, MIL STD 883 D Methode 1011.9, MIL STD 810 Methode 503.1

Main Specification

High Temperature Chamber +10 ... +200°C
Low Temperature Chamber -75 ... +180°C
Temperature Deviation over Time ≤± 1,0°C
≤± 1,5°C in the specimen basket
Temperature Change Specimen transfer time between high and low temperature test chamber <10 sec.
Minimum Recovery Time of Injected Sample <2 minutes (using Nitrogen)
Test-room Dimensions (W x H x D) 620 x 720 x 720 (mm)
Specimen Loading Weight 50 kp distributed on 3 shelves
 

Speed of Temperature Changes

High Temperature Test Chamber During Cooling Process up to 1°C /min in the average
High Temperature Test Chamber During Heating Process up to 5°C /min in the average
Low Temperature Test Chamber During Cooling Process up to 5°C /min in the average
Low Temperature Test Chamber During Heating Process up to 5°C /min in the average